7/6/2007 | Mike McNaughton | The spike at DP West at 6:30 GMT is a calibration. | Station Info | Data Graph |
1/26/2007 | Mike McNaughton | The gamma signal wires were not connected so the gamma data have not been correct since the station was moved in April, 2006. At 23:00 UTC, the connectors were grounded so the gamma reading should be zero from now until we procure and install the correct voltage dividers. | Station Info | Data Graph |
8/11/2003 | Allen Treadaway | A spike, reading 31.8 /hr, occurred on 8/05/03 at approximately 09:30 MDT. There is was no readily apparent cause for this spike and it is still being investigated. | Station Info | Data Graph |
7/22/2003 | Allen Treadaway | The power was restored on 7/21/03 pm and the phone modem was repaired by Qwest on 7/22/03 at about 1600 Z. Some data was restored to the database but not all. During the time the power was low, some data was lost. | Station Info | Data Graph |
7/22/2003 | D Young | At approximatly 19:30 Z on 7/16/03 the DP station went down due to power loss and a phone line that has failed. | Station Info | Data Graph |
6/26/2003 | Allen Treadaway | Calibration was performed on the gamma sensor (PIC) at about 10:45 MDT. | Station Info | Data Graph |
12/16/2002 | Allen Treadaway | At about 23:00Z, a new correction factor for the most recent PIC (pressurized ion chamber) calibration was entered into the software program. This will cause the gamma detector readings to slightly increase by about 7.6% | Station Info | Data Graph |
11/20/2002 | | At about 14:00Z on 11/12/2002, the 12 Vdc battery was found to be reading only about 9.8 Vdc. Upon further inspection at the station, it was found that the AC power recepticle breaker was tripped off. After a reset, the charging of the battery began and voltages quickly came back to the normal 12-13Vdc range. When this happened the pressure sensor began reading normal once more and it appeared as a rapid rise on the data graph at about 19:15-19:45Z on 11/12/2002. | Station Info | Data Graph |
11/1/2002 | Mike McNaughton | The 247.5 microR/h reading at 16:30 Z October 31 was caused by a calibration source. | Station Info | Data Graph |
10/22/2002 | D Young | At 15:55Z Kevin Anderson put a calibration source on the gamma
detector to perform a calibration test. Source will be removed at
16:15Z. | Station Info | Data Graph |
|